EasyManua.ls Logo

Endress+Hauser Proline t-mass 65F - Group OPERATION; LANGUAGE; ACCESS CODE; DEFINE PRIVATE CODE

Endress+Hauser Proline t-mass 65F
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Group OPERATION Device Functions Proline t-mass 65
16 Endress+Hauser
6 Group OPERATION
Function description OPERATION
LANGUAGE Use this function to select the language for all texts, parameters and messages shown on
the local display.
Options:
ENGLISH
DEUTSCH
FRANCAIS
ESPANOL
ITALIANO
NEDERLANDS
NORSK
SVENSKA
SUOMI
PORTUGUES
POLSKI
CESKY
Factory setting:
Country dependent (see Page 71 ff.)
! Note!
If you press the
O S keys at startup, the language defaults to ENGLISH.
ACCESS CODE All data of the measuring system are protected against inadvertent change. Programming
is disabled and the settings cannot be changed until a code is entered in this function.
If you press the
O S key in any function the measuring system automatically goes to this
function and the prompt to enter the code appears on the display (programming disa-
bled).
You can enable programming by entering the personal code
(Factory setting = 65, see the DEFINE PRIVATE CODE function)
User input:
max. 4-digit number: 0 to 9999
! Note!
The programming levels are disabled if you do not press a key within 60 seconds fol-
lowing return to the HOME position.
You can also disable programming in this function by entering any number (other than
the defined private code).
Your Endress+Hauser service representative can be of assistance if you lose your per-
sonal code.
DEFINE PRIVATE CODE Use this function to define a personal code number for enabling programming.
User input:
max. 4-digit number: 0 to 9999
Factory setting:
65
! Note!
If you define the personal code number = 0, the programming is always enabled.
Programming has to be enabled using the factory setting before this code can be
changed.
When programming is disabled the function can’t be changed, this precaution prevents
others from changing your personal code without your knowledge and consent.
STATUS ACCESS Use this function to check the access status for the function matrix.
Display:
ACCESS CUSTOMER (Parameterization enabled)
LOCKED (Parameterization disabled)

Other manuals for Endress+Hauser Proline t-mass 65F

Related product manuals