7 Group USER INTERFACE Device Functions Proline t-mass 65
18 Endress+Hauser
7 Group USER INTERFACE
Function description USER INTERFACE
ASSIGN LINE 1 Use this function to define the display value assigned to the main line (the upper line of
the local display) for display during normal measuring operation.
Options:
OFF
MASS FLOW
MASS FLOW IN %
TEMPERATURE
TOTALIZER 1
TOTALIZER 2
CORRECTED VOLUME FLOW
CORRECTED VOLUME FLOW IN %
ACTUAL CURRENT INPUT
Factory setting:
MASS FLOW
! Note!
The function ACTUAL CURRENT INPUT is only available if a current input module is fit-
ted to the transmitter and is enabled (see function ASSIGN CURRENT INPUT, Page 50)
ASSIGN LINE 2 Use this function to define the display value assigned to the additional line (the bottom
line of the local display) for display during normal measuring operation.
Options:
OFF
MASS FLOW
MASS FLOW IN %
MASS FLOW BARGRAPH IN %
TEMPERATURE
TOTALIZER 1
TOTALIZER 2
TAG NAME
OPERATION/SYSTEM CONDITION
CORRECTED VOLUME FLOW
CORRECTED VOLUME FLOW IN %
CORRECTED VOLUME FLOW BARGRAPH IN %
ACTUAL CURRENT INPUT
Factory setting:
TOTALIZER 1
! Note!
The function ACTUAL CURRENT INPUT is only available if a current input module is fit-
ted to the transmitter and is enabled (see function ASSIGN CURRENT INPUT, Page 50)
100% VALUE LINE 1 Use this function to define the flow value to be shown on the display as the 100% value
of the variable assigned to line 1.
! Note!
This function is not available unless one of the following was selected in the ASSIGN
LINE 1 function:
• MASS FLOW IN %
• CORRECTED VOLUME FLOW IN %
User input:
5-digit floating-point number
Factory setting:
• Country dependent: 10 kg/h or 10 lb/h (if MASS FLOW IN % or MASS FLOW BAR-
GRAPH IN % is selected)
• Country dependent: 10 Nm
3
/h or 10 Sm
3
/h (if CORRECTED VOLUME FLOW IN %
or CORRECTED VOLUME FLOW BARGRAPH IN % is selected)