Device Functions Proline t-mass 65 19 Group SENSOR DATA
Endress+Hauser 65
INTERNAL HEIGHT Use this function to enter the internal height of a rectangular duct for an insertion type
sensor.
! Note!
This function is not available for t-mass F type sensors or if CIRCULAR is selected in the
function PIPE TYPE.
User input:
max. 5-digit floating point number: 75.0 to 2000.0
Factory setting:
150.0 (MILLIMETER) or 6.0 (INCH)
! Note!
• Always use duct internal dimensions
• The appropriate unit is taken from the function UNIT LENGTH in the function group
SYSTEM UNITS (see Page 11).
INTERNAL WIDTH Use this function to enter the internal width of a rectangular duct for an insertion type
sensor.
! Note!
This function is not available for t-mass F type sensors or if CIRCULAR is selected in the
function PIPE TYPE.
User input:
max. 5-digit floating point number: 75.0 to 2000.0
Factory setting:
150.0 (MILLIMETER) or 6.0 (INCH)
! Note!
• Always use duct internal dimensions
• The appropriate unit is taken from the function UNIT LENGTH in the function group
SYSTEM UNITS (see Page 11).
ZERO POINT This function shows the current zero point correction value for the sensor. The zero point
is determined by the ZERO POINT ADJUST function.
Display:
5-digit floating point number including sign:
-20.000 to +20.000
Factory setting:
depends on calibration
FLOW CONDITIONER This function indicates if the flowmeter has been calibrated with or without a flow condi-
tioner.
Display:
WITH
WITHOUT
Factory setting:
WITHOUT
Function description SENSOR DATA