21 Group SIMULATION SYSTEM Device Functions Proline t-mass 65
68 Endress+Hauser
21 Group SIMULATION SYSTEM
Function description SIMULATION SYSTEM
SIMULATION FAILSAFE
MODE
Use this function to set all inputs, outputs and totalizer to their defined fault response
modes, in order to check whether they respond correctly. During this time, the device
status message #691 “SIM. FAILSAFE” appears on the display.
Options:
OFF
ON
Factory setting:
OFF
SIMULATION
MEASURAND
Use this function to set all inputs, outputs and totalizer to their defined flow-response
modes, in order to check whether they respond correctly. During this time, the words
“SIMULATION MEASURAND” appear on the display.
Options:
OFF
MASS FLOW
CORRECTED VOLUME FLOW
TEMPERATURE
Factory setting:
OFF
" Caution!
• The flowmeter cannot be used for measuring while this simulation is in progress.
• The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND
Use this function to define a selectable value (e.g. 12 kg/s). This value is used to test
downstream devices and the flowmeter itself.
! Note!
This function is not available unless the SIMULATION MEASURAND function is active.
User input:
5-digit floating-point number
Factory setting:
0 kg/h; lb/h (MASS FLOW)
0 Nm
3
/h; Sm
3
/h (CORRECTED VOLUME FLOW)
0 °C; +32 °F (TEMPERATURE)
(country dependent, → Page 71 ff.)
" Caution!
The setting is not saved if the power supply fails.