Device Functions Proline t-mass 65 8 Group TOTALIZER 1/2
Endress+Hauser 21
8 Group TOTALIZER 1/2
Function description TOTALIZER 1/2
ASSIGN TOTALIZER Use this function to assign a measured variable to the totalizer.
Options:
OFF
MASS FLOW
CORRECTED VOLUME FLOW
Factory setting:
MASS FLOW
! Note!
The totalizer is reset to “0” as soon as the selection is changed.
SUM Use this function to view the total for the totalizer measured variable aggregated since
measuring commenced. The value is positive.
Display:
max. 7-digit floating-point number, including unit (e.g. 15467.04 kg)
! Note!
The totalizer response to faults is defined in the FAILSAFE MODE function (see
Page 23).
OVERFLOW Use this function to view the overflow for the totalizer aggregated since measuring com-
menced.
Total flow quantity is represented by a floating decimal point number consisting of max.
7 digits. You can use this function to view higher numerical values (>9,999,999) as over-
flows. The effective quantity is thus the total of OVERFLOW plus the value returned by
the SUM function.
Example:
Reading for 2 overflows: 2 E7 kg (= 20,000,000 kg)
The value returned by the SUM function = 196,845.7 kg
Effective total quantity = 20,196,845.7 kg
Display:
Integer with exponent, including unit, e.g. 2 E7 kg