8 Group TOTALIZER 1/2 Device Functions Proline t-mass 65
22 Endress+Hauser
UNIT TOTALIZER Use this function to define the unit for the totalizer measured variable, as selected before-
hand.
Options (for the MASS FLOW assignment):
Metric
→ g; kg; t
US
→ oz; lb; ton
Arbitrary unit
→ _ _ _ _
Factory setting:
kg or lb (country dependent, → Page 71 ff.)
Options (for the CORRECTED VOLUME FLOW assignment):
Metric
→ Nl; Nm
3
US → Sm
3
; Scf
Factory setting:
Nm
3
or Sm
3
(country dependent, → Page 71 ff.)
RESET TOTALIZER Use this function to reset the sum and the overflow of the totalizer to zero (= RESET).
Options:
NO
YES
Factory setting:
NO
! Note!
If the transmitter has a status input and is appropriately configured, a reset for the total-
izer can also be triggered by a pulse (see the ASSIGN STATUS INPUT function, Page 48.
Function description TOTALIZER 1/2