Device Functions Proline t-mass 65 7 Group USER INTERFACE
Endress+Hauser 19
100% VALUE LINE 2 Use this function to define the flow value to be shown on the display as the 100% value
of the variable assigned to line 2.
! Note!
This function is not available unless one of the following was selected in the ASSIGN
LINE 2 function:
•MASS FLOW IN %
CORRECTED VOLUME FLOW IN %
MASS FLOW BARGRAPH IN %
CORRECTED VOLUME FLOW BARGRAPH IN %
User input:
5-digit floating-point number
Factory setting:
Country dependent: 10 kg/h or 10 lb/h (if MASS FLOW IN % or MASS FLOW BAR-
GRAPH IN % is selected)
Country dependent: 10 Nm
3
/h or 10 Sm
3
/h (if CORRECTED VOLUME FLOW IN %
or CORRECTED VOLUME FLOW BARGRAPH IN % is selected)
FORMAT Use this function to define the maximum number of places after the decimal point dis-
played for the reading in the main line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
XX.XXX
! Note!
Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
The places after the decimal point as computed by the flowmeter cannot always be dis-
played, depending on this setting and the engineering unit. In these instances an arrow
appears on the display between the measured value and the engineering unit (e.g. 1.2
kg/h), indicating that the measuring system is computing with more decimal
places than can be shown on the display.
DISPLAY DAMPING Use this function to enter a time constant defining how the display reacts to severely fluc-
tuating flow variables, either very quickly (enter a low time constant) or with damping
(enter a high time constant).
User input:
0 to 100 s
Factory setting:
3 s
! Note!
Setting the time constant to zero seconds switches off the damping function.
CONTRAST LCD Use this function to optimize display contrast to suit local operating conditions.
User input:
10 to 100%
Factory setting:
50%
Function description USER INTERFACE