Device Functions Proline t-mass 65 7 Group USER INTERFACE
Endress+Hauser 19
100% VALUE LINE 2 Use this function to define the flow value to be shown on the display as the 100% value
of the variable assigned to line 2.
! Note!
This function is not available unless one of the following was selected in the ASSIGN
LINE 2 function:
•MASS FLOW IN %
• CORRECTED VOLUME FLOW IN %
• MASS FLOW BARGRAPH IN %
• CORRECTED VOLUME FLOW BARGRAPH IN %
User input:
5-digit floating-point number
Factory setting:
• Country dependent: 10 kg/h or 10 lb/h (if MASS FLOW IN % or MASS FLOW BAR-
GRAPH IN % is selected)
• Country dependent: 10 Nm
3
/h or 10 Sm
3
/h (if CORRECTED VOLUME FLOW IN %
or CORRECTED VOLUME FLOW BARGRAPH IN % is selected)
FORMAT Use this function to define the maximum number of places after the decimal point dis-
played for the reading in the main line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
XX.XXX
! Note!
• Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the flowmeter cannot always be dis-
played, depending on this setting and the engineering unit. In these instances an arrow
appears on the display between the measured value and the engineering unit (e.g. 1.2
→ kg/h), indicating that the measuring system is computing with more decimal
places than can be shown on the display.
DISPLAY DAMPING Use this function to enter a time constant defining how the display reacts to severely fluc-
tuating flow variables, either very quickly (enter a low time constant) or with damping
(enter a high time constant).
User input:
0 to 100 s
Factory setting:
3 s
! Note!
Setting the time constant to zero seconds switches off the damping function.
CONTRAST LCD Use this function to optimize display contrast to suit local operating conditions.
User input:
10 to 100%
Factory setting:
50%
Function description USER INTERFACE