Device Functions Proline t-mass 65 13 Group STATUS INPUT
Endress+Hauser 49
SIMULATION STATUS
INPUT
Use this function to activate simulation of the status input, in other words to trigger the
function assigned to the status input (see the ASSIGN STATUS INPUT function on
Page 40).
Options:
OFF
ON
Factory setting:
OFF
! Note!
The “SIMULATION STATUS INPUT” message indicates that simulation is active.
The flowmeter continues to measure while simulation is in progress, i.e. the current
measured values are output correctly via the outputs.
" Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
STATUS INPUT
Use this function to select the level to be assumed at the status input during the simula-
tion.
! Note!
This function is not available unless the ON setting was selected in the SIMULATION
STATUS INPUT function.
Options:
HIGH
LOW
Factory setting:
LOW
! Note!
If exiting this function and the SIMULATION STATUS INPUT function is active (= ON),
the message END SIMULATION is displayed. Select YES to end simulation or NO to con-
tinue simulation.
" Caution!
The setting is not saved if the power supply fails.
Function description STATUS INPUT