17 Group GAS MIXTURE Device Functions Proline t-mass 65
60 Endress+Hauser
GAS TYPE 3 View Gas Type 3.
! Note!
This function is not available if NUMBER OF GAS COMPONENTS selection is <3
Display:
Gas list the same as GAS TYPE 1
Factory setting:
NOT DEFINED
% FRACTION 3 View % Fraction 3.
! Note!
This function is not available if NUMBER OF GAS COMPONENTS selection is <3
Display:
000.00 % to 100.00 %
Factory setting:
000.00 %
GAS TYPE 4 View Gas Type 4.
! Note!
This function is not available if NUMBER OF GAS COMPONENTS selection is <4
Display:
Gas list the same as GAS TYPE 1
Factory setting:
NOT DEFINED
% FRACTION 4 View % Fraction 4.
! Note!
This function is not available if NUMBER OF GAS COMPONENTS selection is <4
Display:
000.00 % to 100.00 %
Factory setting:
000.00 %
GAS TYPE 5 View Gas Type 5.
! Note!
This function is not available if NUMBER OF GAS COMPONENTS selection is <5
Display:
Gas list the same as GAS TYPE 1
Factory setting:
NOT DEFINED
Function description GAS MIXTURE