EasyManua.ls Logo

Endress+Hauser RIA14 - 6.4 Access to the operating menu via the operating tool; 6.4.1 Configuration via interface & PC configuration software FieldCare Device Setup

Endress+Hauser RIA14
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation options RIA14
16 Endress+Hauser
Menu Operating function Operating function Operating function
Parameter Default/
selection
Parameter Default/
selection
Parameter Default/
selection
Standard dEF 0 to NAMUR
20.5
3.60 NAMUR 3.6
to NAMUR
20.5
3.80
User entry Edit
NAMUR 20.5
2)
N2050
NAMUR 21.0
2)
N2100
Test
TEST
NAMUR
3.80 to
NAMUR
21.0
20.5 20.5 to
25 mA
21.0 Off OFF
Open collect. OUT
Display DISP
Service
SERV
Service code
SCODE
Parameter reset
3)
PRSET
---- Yes YES
No NO
1) Only if DIM = TEXT
2) Only if NAMUR = Edit
3) Can only be performed by service staff
6.4 Access to the operating menu via the operating tool
6.4.1 Configuration via interface & PC configuration software
FieldCare Device Setup
L
WARNING
Loss of explosion protection when housing is open
The device must be configured outside the hazardous area.
During configuration with FieldCare, the device may assume undefined statuses! This may
result in the undefined switching of outputs and relays.
To configure the device using the FieldCare Device Setup software, connect the device to
your PC. You need a special interface adapter for this purpose: the Commubox FXA291.
The 4-pin connector of the interface cable must be plugged into the corresponding socket
in the device and the USB connector must be plugged into a free USB port on the PC.

Table of Contents

Other manuals for Endress+Hauser RIA14

Related product manuals