RIA14 Commissioning
Endress+Hauser 21
Program information
Designation Description
Program name 'PNAME' Displays the name of the software loaded in the device (7-digit)
Display cannot be edited.
Firmware version 'FWVER' Displays the name of the firmware loaded in the device (8-digit)
Display cannot be edited.
Press the + or - key to scroll horizontally through the 7 or 8-digit display values.
Alarm limits (NAMUR)
PARAM → NAMUR
The alarm limits are set to NAMUR values at the factory. These values can be used as
default values (DEF) or edited freely (EDIT).
You can change the following operating items if the 'Edit' operating item has been selected:
Designation Description
NAMUR 3.6 Value range: 0 mA... < Namur 3.8
Default: 3.60
NAMUR 3.8 Value range: Namur 3.6 < x < Namur 20.5
Default: 3.80
NAMUR 20.5 Value range: Namur 3.8 < x < Namur 21.0
Default: 20.50
NAMUR 21.0 Value range: Namur 20.5 < x < Namur 25 mA
Default: 21.00
The Namur limits are listed in ascending order.
Test (TEST)
PARAM → TEST → Options OFF, OUT, DISP
Certain device functions can be tested automatically.
off: OFF (default)
Open Collector: OUT
Display: DISP
7.3.5 Service level (SERV)
This level can only be selected after entering the service code (only available for service
personnel).
Resetting the settings (PRSET)
PRSET - Perform a reset
Service personnel can reset the settings to the default values.
Reset: After YES is selected, the operating parameters are set to the factory default values.
Options: YES or NO
Default: NO
If the default values are set, the option selected is automatically reset to NO.