Diagnosis and troubleshooting RIA14
22 Endress+Hauser
8 Diagnosis and troubleshooting
8.1 General troubleshooting
L
WARNING
Danger! Electric voltage
Do not operate the device in an open condition for device troubleshooting.
8.2 Diagnostic list
Errors that occur during the self-test or during operation are immediately shown on the
display. Error messages that can be acknowledged are deleted after a key is pressed. A
fault has occurred if the hardware for writing and reading data (EEPROM) is defective or if
data cannot be read correctly from the EEPROM.
The errors are defined as:
Error code Meaning
C561 Display overrun
F041 Sensor error (0 mA < input ≤ 2 mA). "Failure" warning symbol is displayed.
F045 Sensor error (2 mA < input ≤ 3.6 mA or input ≥ 21 mA ). "Failure" warning
symbol is displayed.
F101 Range undershoot (input between 3.6 mA and 3.8 mA). "Failure" warning
symbol is displayed.
F102 Range overshoot (input between 20.5 mA and 21.0 mA). "Failure"
warning symbol is displayed.
F261 Error: EEPROM "Failure" warning symbol is displayed.
F282 The parameter data could not be saved. "Failure" warning symbol is
displayed.
F283 Incorrect parameter data "Failure" warning symbol is displayed.
F431 Incorrect reference values "Failure" warning symbol is displayed.
8.2.1 Firmware history
Revision history
The version number on the nameplate and in the Operating Instructions indicates the
device release: XX.YY.ZZ (example 01.02.01).
XX Change to main version
No longer compatible The device and Operating Instructions change.
YY Change to functions and operation
Compatible The Operating Instructions change.
ZZ Fixes and internal changes
No changes to the Operating Instructions
Date Software version Software modification Documentation
08/2009 01.00.00 Original software BA278R/09/en/08.09
09/2010 01.00.zz No change to functions and operation BA278R/09/en/13.10
12/2011 01.00.zz No change to functions and operation BA00278R/09/en/01.11