EasyManuals Logo

Evoqua IONPURE IP-MXM30 User Manual

Evoqua IONPURE IP-MXM30
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
IONPURE
®
MX CEDI Modules
Page 15 of 32 IP-MAN-MX-Rev E
5. START-UP PROCEDURE
5.1. Test Interlocks
Test flow switches and other interlocks, including the RO interlock if applicable
Test pressure relief if applicable
Set alarm points
5.2. Startup of MX Modules
Make sure that modules are correctly connected to the DC power source.
Make sure that the MX product line is directed to drain.
Turn on the feedwater. Adjust valves to obtain the desired flows and pressures in the
product and reject streams. The reject flow is typically set at about 11% of the product
flow (this gives a water recovery of 90%)*. Valves are adjusted so the product outlet
pressure is about 2 to 5 psig (0.1 to 0.3 bar) higher than the reject outlet pressure at
the desired flow rates.
Adjust the DC power supply to the current setting calculated in section 3.4.3
Test all flow switches and interlocks to ensure MX DC power is shut off when flow is
interrupted.
Continue to direct the product water to drain until it reaches the desired quality.
Once product reaches the desired quality, connect to process. Readjust pressures as
required to maintain product outlet pressure 2 to 5 psi (0.1 to 0.3 bar) above the reject
outlet pressure.
When the system is at steady state (quality in specification and stable operation),
record operating data on the data sheet provided at the end of section 4.0.
5.3. Additional Notes
MX recovery of 95% is limited to systems with feed water total hardness of less than
or equal to 0.2 ppm as CaCO
3
. Reduction of hardness is normally accomplished with
ion-exchange softening or with 2 pass RO.
There is a minimum MX reject flow of:
o 2.5 lph for each MX30
o 5.0 lph for each MX60
o 10 lph for each MX125
o 19 lph for each MX250
o 36 lph for each MX500
Therefore, 95% recovery may not be attainable below nominal product flow.
For operation at 95% recovery, feed water silica must be 0.5 ppm as SiO
2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Evoqua IONPURE IP-MXM30 and is the answer not in the manual?

Evoqua IONPURE IP-MXM30 Specifications

General IconGeneral
BrandEvoqua
ModelIONPURE IP-MXM30
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals