IONPURE
®
MX CEDI Modules
Page 30 of 32 IP-MAN-MX-Rev E
8. SHUTDOWN AND STORAGE
This section contains shutdown procedures for a MX module. Under certain
circumstances, bacterial growth can occur quickly in water left stagnant within each
module and the overall system.
8.1 System Shutdown (for periods exceeding 7 days)
Shut off feed water to MX module(s)
Drain standing water out of MX module(s)
Close isolation valves to prevent evaporation of water in membranes and resins.
8.2 Startup After Shutdown
Divert product outlet to drain.
Turn on feed water to MX module(s)
If desired, sanitize MX module(s).
Operate unit with DC power on, flushing to drain.
Send to use point when desired product water quality achieved.
8.3 Disposal
It is not practical to rebuild MX modules
At end-of life they can be disposed of as normal (non-hazardous) waste
o First flush with 5% NaCl for about 5 minutes at minimum to nominal flow rate
o Then flush with tap water for at least 30 minutes
o Both steps with DC power off.