Troubleshooting
108 FIP-400B
Solving Common Problems
An APC fiber is
connected to an
FIP-435B probe, the blue
LED is blinking and the
motor is not running
Try to put the fiber back in place.
Refresh rate is very low
Ensure that the power saving mode is deactivated.
On your smart device, close the applications that are not
in use.
Reduce the number of probes operating in the vicinity.
Bring the probe and the smart device closer together.
The FIP status LED blinks
red for 2 seconds in Live
Video mode and turns
from unlit to blue (auto
focus timeout)
Try to put the fiber back in place.
The FIP status LED blinks
red for 2 seconds when a
capture is performed
and no analysis results
are available
There was an analysis error. Repeat the inspection process.
The list of available FIPs
is empty.
Ensure that the Wi-Fi is on.
Ensure that the probe is on.
A probe is no longer
listed among the list of
available FIPs for
connection.
Wait a few minutes for the probe to appear in the list of
available FIPs.
Connect the Wi-Fi probe with a USB cable.
The ConnectorMax2
Mobile application icon
is not available in the list
of the applications
displayed.
Ensure that the smart device is not in Safe Mode. If so,
restart the smart device.
Ensure that the ConnectorMax2 Mobile application was
installed from the current user account (not from another
account).