3.4 Clinician Sub-menus
The Clinician Menu has four sub-menus: Therapy, Comfort, Device, and Service Info.
Below is a guide to the settings in each sub-menu that you can Activate/Deactivate, or, choose to show/
hide information that appears in the Patient Menu. For more detailed information, please refer to the relevant
section in this Clinician Guide.
Sub-menu Settings
Therapy Options
Auto Mode
CPAP Mode
Therapy Mode
Auto:
Selects auto-adjusting mode and the minimum and maximum pressure
range.
CPAP:
Selects fixed pressure mode and allows you to set the prescribed
pressure.
Auto Mode
Min and Max pressure:
Sets minimum and maximum pressure. These can also be adjusted
independently if Auto Mode has been selected.
CPAP Mode
Set pressure:
Allows you to set the prescribed pressure.
Comfort Options
SensAwake
Activate SensAwake:
Activates or Deactivates the SensAwake feature.
If No
is selected, SensAwake will be Deactivated, you will be taken
back to the main Clinician Menu screen, and the SensAwake Patient
Menu item will be removed from the Patient Menu.
SensAwake pressure:
When SensAwake is Activated, the next automatic menu screen will
allow you to adjust the SensAwake pressure value.
Allow patient to deactivate:
If Yes
is selected, SensAwake will be visible in the Patient Menu
and the patient will be able to turn SensAwake On or O. Patients
cannot adjust the SensAwake pressure.
If No
is selected, SensAwake will disappear from the Patient
Menu so that the patient cannot choose to switch SensAwake On or O.
SensAwake will still run in the background as set up by the clinician.