F&P SleepStyle Product Technical Guide | 67
Clinician Guide – Page 25
A-25
WATER CHAMBER AND CHAMBER SEAL
1. Press the lid latch and open the device lid.
2. Take the water chamber out of the device.
3. Remove the chamber seal from the top of the
water chamber and put aside.
4. Lift the chamber tab and lift the chamber lid to
open.
OUTLET SEAL
1. Grip the outlet seal tab.
2. Gently pull the outlet seal out of the device.
10.2 Cleaning and High-Level
Disinfection of the Device and
Accessories
To maintain optimal therapy and presentation of
the device and accessories, regular cleaning is
recommended. Refer to the cleaning instructions
in section 6 – Caring for Your Device of the F&P
SleepStyle Use and Care Guide.
For reprocessing between patients, follow the
cleaning and then the high-level disinfection
instructions below.
Warnings
To avoid electric shock:
• Do not use bleach, alcohol, or cleaners with
citrus or other natural oils. These substances
may degrade the device and accessories.
• Do not immerse the device in water or any
other liquid.
To avoid incorrect therapy:
• Only clean the device and accessories
according to the cleaning instructions
below.
• Do not clean or disinfect the ThermoSmart
breathing tube with hot water. This may
cause deformation of the tube and reduce
therapeutic pressure.
Cautions
General:
Replace the device and accessories if there is
any sign of cracking, deformation, discoloration
or leaking. It is recommended that you inspect
the device, breathing tube, water chamber,
chamber seal, outlet seal, air filter and elbow, on
a regular basis after cleaning. See section 12 –
Replacement Parts.