Chapter 1: Introduction to the FaroArm
your FaroArm. For information on ordering additional or
replacement battery packs, contact FARO’s Customer Service
800.736.2771 (North America), 407.333.3182 (Worldwide), or
FAX 407.333.8056.
Troubleshooting
Error time-out on USB port
Lost communication, check USB cable and connection.
Check A/C power and On/Off Switch.
Error message on host computer
Contact FARO’s Customer Service 800.736.2771 (North
America), 407.333.3182 (Worldwide), or FAX 407.333.8056.
Have the message ready.
Probe calibration fails
Calibrate the probe again, and check that all points were
digitized properly.
Make sure that probe tip is in contact with hole, or sphere.
Use the other method (sphere or hole).
Buttons (FRONT & BACK) will not respond
Make sure that all of the encoders are referenced. See
“Referencing the Encoders” on page 11.
Turn auxiliary port OFF. See “Edit Probe” on page 31.
Check buttons in Diagnostic command. See “Diagnostics” on
page 41.
Single-point certification shows excessive error
You must use a ball probe. Check that it is installed correctly.
See “Installing Probes” on page 15.
Recalibrate probe. See “Calibrate Probe” on page 33.
During the test watch the probe to ensure flush seating in the
hole. See “Certification” on page 18.