FLUXUS F70x 15 SuperUser Mode
134 UMFLUXUS_F7V4-6-2EN, 2017-10-01
15 SuperUser Mode
The SuperUser mode offers the possibility of an advanced analysis of the signal and the measured values as well as the
definition of additional parameters adapted to the measuring point, in order to achieve better measuring values or during
experimental work. Features of the SuperUser mode are:
Defaults will not be observed.
There are no plausibility checks when parameters are being entered.
There is no check whether the entered parameters are within the limits determined by the laws of physics and technical
data.
The cut-off flow is not active.
A value for the number of sound paths has to be entered.
Some menu items that are not visible in the normal the normal mode are displayed.
15.1 Activation/Deactivation
Press key C. Enter HotCode 071049.
Press key C. Enter HotCode 071049 again to deactivate the SuperUser mode.
15.2 Transducer Parameters
In the SuperUser mode, the menu item Transducer Type will be displayed at the end of the input in the program branch
Parameter, even if the transducers are detected by the transmitter.
15.3 Defining the Flow Parameters
In the SuperUser mode, it is possible to define some flow parameters (profile bounds, correction of the flow velocity) for
the specific application or measuring point.
Attention! The SuperUser mode is intended for experienced users with advanced application knowledge. The
parameters can affect the normal measuring mode and lead to wrong measuring values or to a fail-
ure of the measurement when a new measuring point is set up.
It is displayed that the SuperUser mode is activated. Press ENTER. The main menu will be
displayed.
It is displayed that the SuperUser mode is deactivated. Press ENTER. The main menu will
be displayed.
Attention! Some of the defined parameters are still active after the deactivation of the SuperUser mode.
Press ENTER.
or
Select Special Version to enter the transducer parameters. Press ENTER.
If Special Version has been selected, the transducer parameters have to be entered.
The transducer parameters have to be provided by the transducer manufacturer. Press
ENTER after each input.
Select Special Funct.\SYSTEM settings\Measuring\Calibration. Press EN-
TER.
Select the measuring channel for which the flow parameters are to be defined. Press EN-
TER.
SUPERUSER MODE
*IS ACTIVE NOW*
SUPERUSER MODE
IS PASSIVE NOW
Transducer Type
Q2E-314
Transducer Type
Special Version
Transd. Data 1
35.99
Measuring
Calibration
Calibrat. data
for Channel A:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F705**-NN Series and is the answer not in the manual?

Flexim FLUXUS F705**-NN Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F705**-NN Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals