FLUXUS F70x 5 Selection of the Measuring Point
24 UMFLUXUS_F7V4-6-2EN, 2017-10-01
disturbance source: reducer
inlet: l ≥ 10 D outlet: l ≥ 5 D
disturbance source: pump
inlet: l ≥ 20 D
Tab. 5.2: Recommended distance from disturbance sources;
D – nominal pipe diameter at the measuring point,
l – recommended distance between disturbance source and transducer position