19 Wall Thickness Measurement (Optional) FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 141
When measuring on a rough surface, applying too much of the coupling compound can
result in wrong measured values. A measurement on a very rough surface might be im-
possible (message NO COUPLING will be displayed). In this case, the surface has to be
smoothed.
Fig. 19.2: Acoustic partition boundary
19.3.6 Storing/Transmission of the Wall Thickness
Press ENTER to stop the measurement and to store or transmit the measured value. The
following display appears if a valid wall thickness has been measured and a measured
value transmission is activated.
19.3.7 Stopping the Wall Thickness Measurement
To quit the wall thickness measurement mode, disconnect the wall thickness measure-
ment from the transmitter.
Select yes to store and/or transmit the measured value.
The wall thickness can be transmitted into the current
parameter record.
The pipe material will be replaced by the material used
for the wall thickness measurement.
If the serial transmission of data is activated, the measured
value will be transmitted.
K o p p e l f l ä c h e
P r ü f k o p f
a k u s t i s c h e T r e n n e b e n e
R o h r l ä n g s a c h s e
wall thickness probe
contact surface
acoustic partition boundary
pipe axis
Transfer Data
no >YES<