B Technical Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 203
Shear Wave Transducers (zone 1)
technical type GDP2NW1 GLP2NW1
order code GSP-NA1NL GSP-NA1NL/LC
transducer frequency MHz 2
medium pressure
1
min. extended bar metal pipe: 20
min. bar metal pipe: 30
plastic pipe: 1
inner pipe diameter d
min. extended mm 15
min. recommended mm 20
max. recommended mm 40
max. extended mm 60
pipe wall thickness
min. mm 1.5
max. mm -
material
housing
PEEK with stainless steel cap and
transducer shoe 304 (1.4301)
contact surface PEEK
degree of protection
according to IEC/
EN 60529
IP65
transducer cable
type 1699 1699
length m 4 9
dimensions
length l mm 84
width b mm 40
height h mm 59
dimensional drawing
operating temperature
min. °C -40
max. °C +130
temperature
compensation
x
l
h
b