9 Measurement
9.4 Display of measured values FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
102
9.4.5 Transducer distance
The transducer distance can be displayed during the measurement by scrolling with .
The optimum transducer distance (here: 51.2 mm) will be displayed in parentheses, followed by the entered transducer
distance (here: 50.8 mm).
The optimum transducer distance might change during the measurement (e.g., due to temperature fluctuations).
A deviation from the optimum transducer distance (here: 0.4 mm) will be compensated internally.
R flow profile
information about the flow profile based on the Reynolds number
T fully turbulent flow profile
L fully laminar flow profile
↕ transition range between laminar and turbulent flow
? unknown, cannot be calculated
F flow velocity
comparison of the measured flow velocity with the flow limits of the system
√ OK, the flow velocity is not within the critical range
↑ the flow velocity is above the current limit
↓ the flow velocity is below the current cut-off flow
0 the flow velocity is within the limit range of the measuring method
? unknown, cannot be measured
Fig. 9.4: Display of the transducer distance
Notice!
Never change the transducer distance during the measurement.
Tab. 9.5: Description of the status line
value explanation
L=(51.2) 50.8 mm
54.5 m3/h