7 Connection
FLUXUS G801 7.3 Outputs
71
UMFLUXUS_G801V1-0EN, 2020-06-26
7.2.1 Cable connection
Remove the blind plug to connect the cable to the transmitter.
Prepare the cable with an M20 cable gland. The used cable has to have a wire cross-section of 0.25…2.5 mm².
Push the cable through the cap nut, compression part and basic part of the cable gland.
Insert the cable into the housing of the transmitter.
Screw the sealing ring side of the basic part into the transmitter housing.
Fix the cable gland by screwing the cap nut onto the basic part.
Connect the cable to the terminals of the transmitter.
7.3 Outputs
Connect the output cable to the transmitter (see section 7.2.1, Fig. 7.14 and Tab. 7.9).
Fig. 7.13: Cable gland
1 cap nut
2 compression part
3 basic part
Important!
The max. permissible voltage between the outputs and against PE is 60 V DC (permanent).
Fig. 7.14: Connection of the outputs to the transmitter
FLUXUS *801**-A10****-AA, *801**-A10****-AP
21 3