7 Connection
7.1 Transducers FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
58
The transmitter has 1 housing. The terminals for the connection of transducers, outputs and power supply are on the rear
side of the housing.
7.1 Transducers
The transmitter has 2 cable glands for the connection of the transducers. If the transmitter has only one measuring
channel, one of the openings is closed with a blind plug.
It is recommended to run the cables from the measuring point to the transmitter before connecting the transducers to
avoid load on the connectors.
Fig. 7.1: Connections of the transmitter
1 – transducers (channel A)
2 – transducers (channel B)
3 – power supply
4 – outputs
5 – equipotential bonding terminal
Notice!
If transducers are replaced or added, the transducer module has to be replaced or added as well.
Fig. 7.2: Connection of the transducers to the transmitter FLUXUS *801**-A10****-AA, *801**-A10****-AP
1 – transducers (measuring channels A and B)
1
2
3
4
5
1