13 Outputs
FLUXUS G801 13.1 Installation of a binary output
115
UMFLUXUS_G801V1-0EN, 2020-06-26
Assignment of a source item
One source item has to be assigned to each selected output.
• Select the physical quantity (source item) to be transmitted to the binary output by the source channel.
• Press ENTER.
The source items and their scroll lists are summarized in the following table.
Function test of the binary output
The function of the output can now be tested.
• Connect a measuring instrument to the terminals of the installed output.
• Select Reed-Relay OFF or Open collect OFF in the scroll list Output Test to test the de-energized state of the
output.
• Press ENTER.
• Measure the resistance at the output. The value has to be high ohmic.
• Select yes to repeat the test, no to return to the menu item SYSTEM settings.
• Press ENTER.
• Select Reed-Relay ON or Open collect. ON in the scroll list Output Test to test the energized state of the output.
• Press ENTER.
• Measure the resistance at the output. The value has to be low ohmic.
• Select yes to repeat the test, no to return to the menu item SYSTEM settings.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\Source item
Tab. 13.1: Configuration of the binary outputs
source item list item output
Limit R1 limit message (Alarm Output R1)
R2 limit message (Alarm Output R2)
R3 limit message (Alarm Output R3)
Impuls from abs(x) pulse without sign consideration
from x > 0 pulse for positive measured values of the volumetric flow rate
from x < 0 pulse for negative measured values for the volumetric flow rate
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\Reed-Relay OFF
Special Funct.\...\B1 Output Test\B1=ON\Again?
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\Reed-Relay ON
Special Funct.\...\B1 Output Test\B1=ON\Again?