13 Outputs
FLUXUS G801 13.2 Configuration of a frequency output as pulse output
117
UMFLUXUS_G801V1-0EN, 2020-06-26
Activation of a frequency output
Select the channel for which the output is to be activated in the program branch Output Options.
Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
Press ENTER until Frequency Output is displayed.
Select yes to activate the output.
Press ENTER.
Enter the number of pulses that is to be assigned to the unit of measurement of the totalizer.
Press ENTER.
Example: 1000 pulses correspond to 1 m³ of the totalized fluid.
The max. flow depending on the upper limit of the frequency and of the pulse value is indicated.
Press ENTER.
Output Options\for Channel A
Output Options\...\Frequency Output
Output Options\...\Frequency Output\Pulses per unit
Output Options\...\Frequency Output\INFO: max flow= 3600.0 m3/h