13 Outputs
FLUXUS G801 13.2 Configuration of a frequency output as pulse output
117
UMFLUXUS_G801V1-0EN, 2020-06-26
Activation of a frequency output
• Select the channel for which the output is to be activated in the program branch Output Options.
• Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
• Press ENTER until Frequency Output is displayed.
• Select yes to activate the output.
• Press ENTER.
• Enter the number of pulses that is to be assigned to the unit of measurement of the totalizer.
• Press ENTER.
Example: 1000 pulses correspond to 1 m³ of the totalized fluid.
The max. flow depending on the upper limit of the frequency and of the pulse value is indicated.
• Press ENTER.
Output Options\for Channel A
Output Options\...\Frequency Output
Output Options\...\Frequency Output\Pulses per unit
Output Options\...\Frequency Output\INFO: max flow= 3600.0 m3/h