7 Connection
FLUXUS G801 7.1 Transducers
69
UMFLUXUS_G801V1-0EN, 2020-06-26
7.1.5 Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced or added, the transducer module has to be replaced or added as well.
Notice!
The serial numbers of the transducer module and the transducer have to be identical. A wrong or incorrectly
connected transducer module will lead to incorrect measured values or to a measurement failure.
Fig. 7.11: Connection of the transducer module – transmitter FLUXUS *801
1 transducer module (measuring channel A)
2 transducer module (measuring channel B)
1 2