5100
Series
Table
1-4.
Current Specifications
DC
Current
Range
Resolution
Compliance
Voltage
Accuracy
(6
months)
(20*0
to 30®
C)
Ripple
and Noise
±(0.2A to
1.99999A)
±(20
mA to
199.999
mA)
±(2
mA to
19.9999
mA)
±(0.2
mA to 1.99999
mA)
±(10
pA’
to
199.999
pA)
10
p
A
1
pA
100
nA
10
nA
1
nA
0
to 2.1V min
0
to
10V min
0 to 10V
min
0 1o 10V
min
0
to
10V mm
±{0.025%
of
setting +0.0025%
of range±Q-01
pA)
Compliance voltage:
>1V
add 0.002%
setting/volt
(0.05%
of output
+0.01
pA)
rms
Measured
with
a
band-
width of 10 Hz to
10
kHz
including random
spikes
(1!
10%
lower
current
available
using
the
Edit Control.
Temperature
Coefficient
(Above 30°C
and Below
20°)
The accuracy
limits increase by
±{10
ppm
of setting
+
2
ppm of rangejrC
Transient Recovery Time
1
Second to
settle to within 0.01%
of
final value
following
any
change in
current
or
compliance
voltage.
Short Term
Stability (10
Minutes)
At any
fixed temperature
from
0°C to
50®
C
the
short
term
stability is
±(50
ppm of setting
+
5
ppm of
range
+
0.002
//A).
Load Regulation
±20
ppm/volt for a change in the output voltage from
t volt to maximum rated
compliance voltage.
Overvoltage
Protection
On
all ranges voitage is limited to not more than 2V great-
er
than maximum rated compliance
voltage due to an
open circuit condition.
The
operator
is alerted
by a
flash-
ing "O.L." on the central display. After
approximately
2
seconds the
calibrator
goes
to
standby.
Guard
The
DC
current section is
guarded
and
a
front panel
terminal labeled
“I
GUARD"
is
provided.
AC
Current
Range
Resolution
Compliance
Voltage
Accuracy
(6
months)
(20°C
to
3Q°C)
Frequency
Total Harmonic
Distortion
and Noise
0.2A
to
1 ,99999
A
20 mA to 199.999
mA
2 mA to
19.9999
mA
0.2
mA to
1.99999
mA
10
pA
'
to
199.999
pA
10
pA
1
pA
100
nA
10 nA
1 nA
0
to 1.4V rms
min
0 to 7V rms min
0 to 7V
rms min
0 to 7V
rms
min
0 to 7V
rms min
±{0.07%
of
setting
+
0 .
01
% ct
range
+2 pA)
Compliance voltage:
>1V
rms
add 0.0G5%
of
setting/volt
50 Hz
to
1 kHz
Accuracy:
±3%
Resolution
1
MSD
Distortion, line
interference
+
noise
including
random spikes
(0.05% of output
+2
pA) rms
{1)
10%
lower current
available
using the Edit Control.
Temperature Coefficient
(Above
3Q°C and Below
20°C)
CURRENT:
Accuracy
limits
increase
by
±(25
ppm
of
setting+10
ppm
of range)/°C.
FREQUENCY:
Accuracy limits increase by
±0.1%/
<I
C-
Load
Regulation
±50
ppm/volt
for
a
change
in the
output voltage
from
1 V to
maximum rated compliance voitage. Load regulation is
met
with reactive
loads with power factors between 0.9
and 1.0.
Transient
Recovery
Time
4
Seconds
to
settle
within
0.02%
for current and
within
0.3% for
frequency
following
any
change
in output cur-
rent, voitage
or
frequency.
Short
Term
Stability
(10
Minutes)
At
any fixed
temperature
from 0
°
C to
50°
C
the short term
stability is
±{0.01 4%
of
setting+0.002%
of range
+0.4pA).
Overvoltage
Protection
On all
ranges
voltage is
limited
to
not
more
than 2V peak
greater
than maximum
rated
compliance
voltage due to
an open circuit condition. The operator
is
alerted by a
flashing "O.L.” on the central display. After approximately
2
seconds
the calibrator
goes
to
standby.
Guard
The AC
current section
is
guarded and a front pane! ter-
minal labeled
“I
GUARD" is
provided.