CALIBRATION ADJUSTMENT PROCEDURE
6-13
S24. External triggering
Purpose; calibrate tTie 0.2V external trigger level.
Calibration equipment:
Fluke 5100B Calibrator
Calibration setup:
BMC
(C
9ANANA
(female)
(n^)
ScopeUdfer
Procedure:
A
-
Apply
a 50
kHz Sine wave signal
with an amplitude of 1
V
peak-to-peak to channel A and also to
the
banana connectors. Use a coaxial signal splitter and
a
BNC(female)’tO'banana(male}
converter
(see calbration setup). (Set the Fluke 51006
to 0.35355V RMS, 5 kHz sine wave).
B
Press the READY softkey.
S25. Random sampling
Purpose: calibration of the random sampling levels.
Calibration equipment:
Tektronix
PG 506 Square Wave Calibration Generator