5-14
CAUBRATION ADJUSTMENT PROCEDURE
Calibration setup:
Procedure:
A
Apply a 1 MHz square wave signal with an amplitude of approximately 600 mV peak-to-peak to
both channel A and B. Setthe generator to the FAST RISE position. Use 50Q terminations.
B
-
Press the READY softkey.
C
Now press the SCOPE softkey to go back to tfie CALIBRATE menu.