CALIBRATION ADJUSTMENT PROCEDURE 5-17
Tabi9 S.3 CaJibrat/on signals for stop
M4...M7.
Calibration Step
Number C^ibration Voltage
300 mV DC
3V DC
30V DC
M7 30OV DC
WARNING: After
you have performed calibration M7, deactivate the Fluke 5100B to remove
the 300V DC.
Always set the Fluke S1008 to 300 mV DC before touching the
connection
cables!
M8/9. External input, 300 mV/3V range: gain calibration
Purpose: calibration of the external input gain in the 300 mV and 3V ranges.
Calibration equipment:
Fluke 51 006 Calibrator
Calibratjon
setup:
Procedure:
A
Apply 300 mV DC to the banana connectors.
B
-
Press
the READY softkey.
C
-
Apply 3V DC to the banana connectors.
D
-
Press the READY softkey.