5-18
CALIBRATION ADJUSTMENT
PROCEDURE
M10. AW ranges Oo calibration
Purpose: callbratior>
of the 0^2 points In all ranges
Calibration
equipment:
none
Calibration
setup:
SHORT CIROJd
A
J i
aaaaa
DC3CIO
oaonn
Q a a
OOeDD
irni
1
D
nop
ScopeMeler
Procedure:
A
Short circuit the banana
connectors.
B
Press the READY
softkey.
M11/12/1
3/14/1 5/1 6. Calibration of the Ohm ranges
Purpose: calibration of the
300£l
3 kO, 30 kO. 300 kCl 3 MO, and 30 ranges.
Calibration equipment:
Fluke 5100B Calibrator