CALIBRATION ADJUSTMENT PROCEDURE
5-19
Calibration setup:
Scopef^er
Procedure:
A
-
Connect lOOnto
banana connectors.
B
-
Press the READY softkey,
C
-
Change the resistance according
to table 5.4. After each calibration press the
READY softkey.
Use the adjust^select keys toadvance/go
back
in
the list.
Tdble
5.4 Calibration signals for step
Calibration
Step Number Calibration Resistance
Mil
100&2
M12
1 kn
M13
10 Id^
M14 100 kn
M15
1 m
M16
10MQ
M17. Voltage ramp calibration
Purpose: calibration of
the voltage ramp of the circuit tester,
Calibration equipment:
none