5-20
CALIBRATION ADJUSTMENT
PROCEDURE
Calibration
aotup:
Procedure:
A
Connect the channel
A
BNC to the red GENERATOR
OUT banana connector, by means of a
BNC cable and
a
BNC
(lemalehto- banana
(male)
connector.
B
Press the READY softkey.
M18. Current ramp calibration
Purpose: calibrate the
current ramp of the circuit tester.
Calibration
equipment:
Fluke
5)006
Cahbrator
Calibration setup:
Procedure:
A
Connect a
resistance
of 100Q
between
both banana connectors. Connect channel A to the red
banana connector. Do
not use a prc^l
Refer
to the
calibration
setup.
B
Press the READY softkey.