CAUBRATION
ADJUSTMENT PROCEDURE
5-23
Proe^ure:
A
-
Connect the channel A BNC to the
red GENERATOR OUT banana
connector, by means o1 a
BNC cable
and a BNC (femate)-to-
banana(male) connector.
B
-
Press
tfie READY softkey.
If all
connections are good, the
SoopeMeter will display the text:
"DC PROBE calibration
In progress".
After
a few seconds the
SoopeMeter will display:
"PROBE successfully calibrated"
and will also
beep once. Now you
can go to the next calibration step.
Connect the channel B BNC to the
red GENERATOR OUT banana
connector,
by means of a BNC cable and
a BNC(femaleHo-banana(male)
connector.
Preoa
the READY softkey. If all connections
are good, the SoopeMeter will
display the text:
"DC PROBE calibration In progress".
After a few seconds the
SoopeMeter will display:
"PROBE
successfully calbrated"
and It will also beep once.
Calibration
is now complete.