CORRECTIVE MAINTENANCE
3
Seop6
mode: not
rnorethan
5 measurements You have
attempted to switch
on
more
(SCOPE mode)
than
five cursor measurements
dlmultaneously
In the cursor function
pop*up
menu.
Solution: Turn off
another cursor function.
Unknown probe or
wrong connection.
No probe or a defective
probe has
(SCOPE/METER mode)
been connected during probe
DC calibration.
Solution: Connect a correct
probe and do another DC calibration. If
the warning Is still displayed,
refer to the troubleshooting
information of the Analog
A2 PC6 later In this section.
No valid
memory setup that
can be used. You have
tried to
recall
a waveform
and
( SCOPE mode
)
the
correspondIng setu
p
(Setu
p
recal I
active), while a setup has
not been saved for
the stored wavefonrt.
Solution:
Choose a waveform for which there is a valid
setup
stored,
or switch off the
"Setup recall*
furtction.
Not executed:
no t2V programming
voltage The CALIBRATE
ScopeMeter softkey
has
(SCOPE/METER
mode)
Deen pressed In
the SERVICE menu
without
the
12V programming voltage being
connected
to the programming
contacts in
the
battery compartment.
Solution: Connect
the 12V programming voltage, before pressing
the CALIBRATE softkey,
NOTE: Cslibration
/s to be done by qusl/fied
eervice personnei Incorrect
calibration data is stored
if
12V
programming
voltage is connected,
while the CALIBRATE
ScopeMeter mode is turned
on.
For
calibration
of the Scopemeter refer
to chapters: ’'Calibration
Adjustment Procedure".
CAL STORE
error: no 12V or no space
left 1. No 12V.
The 12V programming voltage
(SCOPE/METER
mode)
xh&t Is connected
to the programming
contacts
in the battery compartment
has
disappeared
during the calibration
adjustments.
2. No space left. The internal
Flash ROMs
with the calibration constants
are full.
Solution: 1
.
Check
the 12V programming voltage
connection in the battery
compartment.
2- The calibration constants
part in the Flash Roms must
be emptied before
other
calibrations
can be made. For refreshing the Flash
ROMs, contact your nearest
Fluke/Philips
Service Center.
PRINTER error: please
reset printer.
No printing or the printing has
stopped via
(SCOPE/METER mode,
modei 97 only)
the optically Isolated RS-
232-C interface
PM9080.
Solution: Check
the settings on the printer
(ON LINE and BAUD RATE). Reset
the printer. Verify
that if
the optically isolated
RS-232'C interface is still connected
to the ScopeMeter.