CORRECTIVE MAINTENANCE
CALIBRATION error wrong input signal(s)
(SCOPE/METEfl mode)
The SoopeMeter has rejected the connected
calibration adjustment signal during
calibration.
Solution: Check the calibration signal and repeat the calibration step. If the signal is correct and the
error message remains, refer to the troubleshooting information of the Analog
A2
PCB
later in this section.
PROBE successfully calibrated.
(SCOPE/METER mode)
The probe calibration has been successful.
**
ERROR
***
PLEASE RESET INSTRUMENT
***
General error message: eomethlng has gone
(SCOPE/METER mode) wrong, which cannot be undone easily.
Solution: Switch off the ScopeMeler and switch lion again, using MASTER RESET: Press the LCD
key and keep it pressed. Now press the ON/OFF key. The ScopeMeler wt\ give two beeps
and will start up in a default condition.
No AUTOSET on time oratt: no channels
(SCOPE mode)
You have tried to do an AUTOSET, while
both channels A and B were switched off
(only waveforms in memory displayed!).
Solution; Switch on channel A and/or channel B before you activate AUTOSET.
REF differs from present meter mode.
(METER mode)
The settings of the ScopeMeter have been
changed, so that previously deteimined
references are not valid.
Solution: Set new
references.
PROBE CAL
Use
AUTO SET to exit.
(SCOPE/METER
mode)
The ScopeMeter has
been set Into the
AC ADJUST
mode for
channel
A or B in the
PROBE
CAL popup menu,
Solution; AC adjust the probe and/or press the AUTO SET key.
AUTO SET .. AUTO SET .. AUTO SET
(SCOPE/METER mode)
The ScopeMeter performs an auto set after
the AUTO SET key has been pressed.
Solution; Wait until the warning dis^pears (about 1 second). If the warning stays, refer to the
troubleshooting information tater In this section.
Connect PROBE to GENERATOR OUT.
(SCOPE/METER mode)
The AC ADJUST or
the DC CAL
item
In
the PROBE CAL pop up menu has been
selected.
Solution: Connect
a
probe
to
the generator
output and
select
AC ADJUST or DC CAL or watt for
five seconds,
DC PROBE calibration in progress
(SCOPE/METER mode)
The DC CAL item in the PROBE CAL
pop-up menu has been selected.
Solution; Wait until the warning disappears.
A
beep signals the end of the DC PROBE calibration.
If
the calibration
has been successful, the message
TROBE successfully calibrated"
will
appear.