PERFORMANCE VERIFICATION PROCEDURE
4-13
Requirements:
TaOle 4.3 Requirements for voltage accuracy test channel A. METER mode.
input signal
^
Requirements
300 mV DC
300 mV RMS AC. 1 kHz
3V DC
3V RMS AC, 1 kHz
30V DC
30V
RMS
AC,
1 kHz
298-0... 302.0V DC
292.5..
. 307.5V RMS AC
2.980..
.3.020V DC
2.925.-.3.075V RMS AC
29-80.-,
3O.20V
DC
29.25..
.30.75V RMS AC
2. DC mV accuracy METER mode
These tests check the accuracy of the DC mV function. The signal must be
supplied 1o the banana input connectors
of
the
ScopaMeter.
Test equipment;
Fluke 5 1006 Catibralor
Test setup:
Seop«Meter
Procedure/requI rementa:
A Apply 300 mV DC to the banana connectors of the ScopeMeter.
B Verify that the readout is between 298 .2... 30 1
.8
mV DC.
C
Apply 3V DC
to
the banana connectors of the
ScopeMeter.
D Verify that the readout is between 2.982...3.0 18V DC.
3. Resistance accuracy
3
^
These tests check the accuracy of the resistance measurement function, The
signal
has to be supplied to the
banar^a
input connectors of the ScopeMeter.