PERFORMANCE VERIFICATION PROCEDURE
4-15
Procedure/requlrementa for diode accuracy teatt
A Set the Fluke 5100B to 1 kn.
B Che<3< that the readout is between 0.420...0.569V DC.
5.
Signal display
and frequency measurement
This test checks the waveform display and the frequency measurement function
in METER MODE.
Test equipment:
Tektronix TG 501 Time Mark Generator
Test setup:
Procedure/requirements for testing waveform display and frequency function:
A
Apply a
1 ms
(IV
peak-to-peak) time
marker signal
to channel A. Use a 50Q termtnatbn.
B Check that a stable (triggered) signal is displayed.
C Check that the frequency
displayed
is between 993...1007 Hz.