4-26
PERFORMANCE VERIFICATION PROCEDURE
L The ScopeMeter display will
look like figure 4.1 3. The generator
must produce a sine wave
signal with an
amplitude of IV and a frequency of 976 Hz (typical values).
A200mVAC
8 2V OFF PROBE 10:1
CSV;
1.CO
V
dt:
2.64
ms
FREQ:
97S
Hz
START
PRINT
9ni79
Figure 4. t3 Generafor produces sine wave signal
500ya/DIV
TRIG.
A/
GENERATE MEASURE PRINT PRINTER
^ ^
FORMAT^ SETUPS
11.
Component test function
»**
Model 97 only!
*»•
This test checks the component test function (slow ramp
voltage and slow ramp current).
Test equipment:
Red
scope probe (delivered with the ScopeMeter)
Test setup:
aius
Settinga/procedure/requiremente;
A Switch on the ScopeMeter
and press the SPECIAL FUNCT key to enter the SPECIAL
FUNCT
menu.
B Now press the MEASURE softkey. This
will reveal the MEASURE pop-up menu.