MPS450
Operators Manual
4-18
Table 4-29. Numeric Codes for BP Channel 3: Static-Pressure Levels Actions (cont.)
Numeric
Control Code
Action
Remote Control
Entry
364 Set BP3 to 60 mmHg static. P3S60
365 Set BP3 to 80 mmHg static. P3S80
366 Set BP3 to 100 mmHg static. P3S100
BP Channel 4: Static-Pressure Levels
Table 4-30 lists the numeric control codes for the BP channel 4: static-pressure levels
functions.
Table 4-30. Numeric Codes for BP Channel 4: Static-Pressure Levels Actions
Numeric
Control Code
Action
Remote Control
Entry
369 Set BP4 to -5 mmHg static. P4S-5
370 Set BP4 to zero. P4S0
371 Set BP4 to 20 mmHg static. P4S20
372 Set BP4 to 40 mmHg static. P4S40
373 Set BP4 to 60 mmHg static. P4S60
374 Set BP4 to 80 mmHg static. P4S80
375 Set BP4 to 100 mmHg static. P4S100
BP Channel 1: Dynamic Waveforms
Table 4-31 lists the numeric control codes for the BP channel 1: dynamic waveforms
functions.
Table 4-31. Numeric Codes for BP Channel 1: Dynamic Waveforms Actions
Numeric
Control Code
Action
Remote Control
Entry
060 Run BP1 arterial dynamic wave 120/80. P1ART
061 Run BP1 radial art dynamic wave 120/80. P1RART
062 Run BP1 left vent dynamic wave 120/0. P1LV
063 Run BP1 right vent dynamic wave 25/0. P1RV
BP Channel 2: Dynamic Waveforms
Table 4-32 lists the numeric control codes for the BP channel 2: dynamic waveforms
functions.