Respiration 3
3-5
2. To adjust the respiration rate, select DOWN or UP. The rate identified on the LCD
screen is active when displayed and remains active until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-rate
setting according to Table 3-3, and select RUN:
Table 3-3. Numeric Codes for Respiration-Rate Settings
Respiration-Rate Setting Numeric Code
0 BrPM 156
15 BrPM 157
20 BrPM 158
30 BrPM 159
40 BrPM 160
60 BrPM 161
80 BrPM 162
100 BrPM 163
120 BrPM 164
Adjusting the Respiration Amplitude (Impedance Variation)
The MPS450 offers four preconfigured settings (ohms) for respiration amplitude: 0.2,
0.5, 1.0, or 3.0.
Action in the Menu-Control Mode
1. Press the top-menu key labeled RESP. If the respiration-amplitude setting has not
been changed, the screen displays the default setting (1.0 OHMS).
2. To adjust the respiration amplitude, select AMPL repeatedly to scroll to the desired
setting. The amplitude identified on the LCD screen is active when displayed and
remains active until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-
amplitude setting according to Table 3-4, and select RUN:
Table 3-4. Numeric Codes for Respiration-Amplitude Settings
Respiration Amplitude Setting Numeric Code
0.2 ohms 195
0.5 ohms 196
1.0 ohms 197
3.0 ohms 198