Cardiac Functions
Fetal / Maternal ECG (Option) 2
2-41
UNIF ACC (uniform acceleration).
5. Select RUN. On the LCD screen, RUNNING flashes while the simulation is active.
6. A manually generated IUP ends after one contraction. To repeat the simulation, select
RUN again.
7. Periodically generated IUPs repeat at the selected period. To terminate a periodic-
IUP simulation, select END.
Alternatively, in the Numeric-Control Mode, press the number keys for an IUP-wave
simulation using one of the codes in Table 2-33, and select RUN. (The selected IUP is
generated according to the specific period that was last selected, whether manual, or once
every 2, 3, or 5 minutes.)
Table 2-33. Numeric Codes for IUP Wave Settings
IUP Wave Setting Numeric Code
IUP OFF 400
IUP UDEC (uniform deceleration) 401
IUP EDEC (early deceleration) 402
IUP LDEC (late deceleration) 403
IUP UACC (uniform acceleration) 404
To select a contraction period in the Numeric-Control Mode, press the number keys for
one of the settings found in Table 2-34, and select RUN. (The selected period is applied
to a run of the specific IUP-wave simulation that was last selected, whether uniform
deceleration, early deceleration, late deceleration, or uniform acceleration.)
Table 2-34. Numeric Codes for IUP Contraction Period Settings
IUP Contraction Period Setting Numeric Code
IUP MANUAL 405
IUP 2 MIN 406
IUP 3 MIN 407
IUP 5 MIN 408