Remote Operations
Codes and Actions 4
4-21
BP Channel 2: Respiration Artifact
Table 4-36 lists the numeric control codes for the BP channel 2: respiration artifact
functions.
Table 4-36. Numeric Codes for BP Channel 2: Respiration Artifact Actions
Numeric
Control Code
Action
Remote Control
Entry
358 Set BP2 respiration artifact off. P2AOFF
359 Set BP2 respiration artifact on. P2AON
BP Channel 3: Respiration Artifact
Table 4-37 lists the numeric control codes for the BP channel 3: respiration artifact
functions.
Table 4-37. Numeric Codes for BP Channel 3: Respiration Artifact Actions
Numeric
Control Code
Action
Remote Control
Entry
367 Set BP3 respiration artifact off. P3AOFF
368 Set BP3 respiration artifact on. P3AON
BP Channel 4: Respiration Artifact
Table 4-38 lists the numeric control codes for the BP channel 4: respiration artifact
functions.
Table 4-38. Numeric Codes for BP Channel 4: Respiration Artifact Actions
Numeric
Control Code
Action
Remote Control
Entry
376 Set BP4 respiration artifact off. P4AOFF
377 Set BP4 respiration artifact on. P4AON
Other Functions
The Other Functions category consists of Temperature, Cardiac-Output Wave/Injectate,
FHR Rate (Fixed), Intrauterine-Pressure (-) Wave, Intrauterine-Pressure (IUP) Period,
View Angle, and Beeper.
Temperature
Table 4-39 lists the numeric control codes for the temperature functions.
Table 4-39. Numeric Codes for Temperature Actions
Numeric
Control Code
Action
Remote Control
Entry
189 Set temperature to 0 degrees C. T0
190 Set temperature to 24 degrees C. T24
191 Set temperature to 37 degrees C. T37
192 Set temperature to 40 degrees C. T40