EasyManua.ls Logo

Fluke CableAnalyzer DSP-4000 Series - Page 173

Fluke CableAnalyzer DSP-4000 Series
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Specifications
Specifications
8
8-13
Typical Measurement Accuracy
The DSP-4000 and DSP-4100 test tools substantially exceed the measurement
accuracy requirements of draft Level III TIA and IEC proposals. Typical
measurement accuracy is determined as follows:
For each performance parameter (for example, source/load return loss, residual
NEXT, etc.) at every frequency, the worst result for any wire pair or wire-pair
combination is obtained.
The values of the worst case wire pair or wire pair combinations are
substituted in the accuracy equation given by TIA and IEC guidelines.
Worst case assumptions for link performance that affect measurement
accuracy are used. These include the assumptions for link return loss,
common-mode-to-differential-mode gain, and differential-to-common-mode
gain.
Figure 8-2 shows typical baseline measurement accuracy at TIA Cat 6/ISO/IEC
Class E channel limits. The values are calculated using the worst case for each
performance parameter specified for field cable testers.
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
0 50 100 150 200 250
Frequency in MHz
Accuracy in dB
Insertion loss
NEXT
ELFEXT
Return Loss
oy83f.eps
Figure 8-2. Typical Measurement Accuracy

Table of Contents

Other manuals for Fluke CableAnalyzer DSP-4000 Series

Related product manuals