Getting Started
Read First: Safety and Operational Information
2
2-3
Never attempt to insert any connector other than an
RJ45 connector into the RJ45 jack. Inserting other
connectors, such as RJ11 (telephone) connectors, can
permanently damage the jack.
Never attempt to send data from a PC to the test tool
while running a cable test. Doing so might cause
erroneous test results.
Never operate portable transmitting devices during a
cable test. Doing so might cause erroneous test
results.
When using the optional channel/traffic link interface
adapter (DSP-LIA013), never run tests with cables
connected to both the cable jack and the monitor jack.
Doing so might cause erroneous test results.
To ensure maximum accuracy of test results, perform
the self-calibration procedure as described in
“Calibrating the Test Tool” in Chapter 6 every 30 days.
To avoid false test results, recharge the battery as
soon as the low battery message appears.