EasyManua.ls Logo

Frontier EGA/PY-3030D - Page 40

Frontier EGA/PY-3030D
83 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-11
Ver.1.36
(11) When the GC sends a READY signal, the dialog box shown in Fig. 4.12 will be displayed.
Manually press the sample cup drop button at the top of the sampler. After the sample cup drops
into the furnace, immediately click the Start button on the dialog box or hit ENTER key on the
PC keyboard. This will start the GC.
This completes the Double-Shot operational sequence. Fig. 4.13 below shows the schematic diagram
for a series of operational events as described above.
Fig. 4.12 Pyrolysis start dialog box in Double-Shot analysis
Fig. 4.13 Sequence program in Double-Shot Analysis
1st TD Initial (min)
1st TD Final (min)
PY
1st TD Initial (°C)
1st TD Final (°C)
Pyrolysis (°C)
1st TD Rate
(°C/min)
1st TD
PY Initial (min)
START
END
A dialog box (Prepare for pyrolysis) is displayed.
If pyrolysis is to be done, pull up the Double-Shot
sampler all the way, then click Yes button. This
will start GC analysis.
When GC is ready, a dialog box (Ready for
pyrolysis) is displayed.
Drop the sample cup. Then immediately click
Start button. This will start GC analysis.
A dialog box (Ready for desorption) is displayed.
Drop the sample cup into the furnace, and immediately click
START button. At this point, the GC analysis is not started.
2nd TD Final (°C)
2nd TD Final (min)
2nd TD Rate
(°C/min)
2nd TD
GC analysis
GC analysis
analysis
GC analysis

Table of Contents