Changing the MIC TRAC with Force-Lok Measuring Force
Materials Needed:
¾ MIC TRAC MT-3000 base unit
¾ Flat face anvils mounted (TF-1F Blocks)
¾
Digital readout
1. To change the measuring force press to
locate Measuring Force key on page 2 of the
display.
2. Press the softkey below Measuring Force to
display the Measurement Force screen. Use
the Increase or Decrease softkeys to increase
or decrease the measurement force. Force is
adjustable in .5 lb. increments with a range of
1.0 to 5.0 lbs.
3. Press to complete the operation and
save the selected value.
Note: The unit must be “re-zeroed” by bringing the
anvils together and applying slight pressure
until the lights up. The display will
read 0.00000 and the gaging force will be
set to the new pressure.
This new pressure setting will remain the
default setting until it is changed, even if the
power is turned off.
MEASUREMENT FORCE
MEAS. FORCE
Select the desired
measurement force from 1 to 5
pounds, with the Softkeys
INCREASE and DECREASE.
2.0S
DECREASE
INCREASE

Other manuals for GAGEMAKER MIC TRAC MT-3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GAGEMAKER MIC TRAC MT-3000 and is the answer not in the manual?

GAGEMAKER MIC TRAC MT-3000 Specifications

General IconGeneral
BrandGAGEMAKER
ModelMIC TRAC MT-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals