Note:
The H or L indication at the left bottom of the dis-
play indicates which operational frequency has been 
selected for measurement, either frequency for the 
12.7 mm probe. 500 kHz (H) only may be used for the 
8 mm probe.
3.4  Entering probe code data
E.g. 12.7 mm H/L probe, code number is 6547. The 
code is stored in the format below, there are code slots 
for three large and three small diameter probes, and 
their selection is sequential.
6 5 4 7
and is equivalent to data stores
D1 D2 D3 D4
Entering probe code data
Step Key Display
1
From “Off” select mea-
surement mode.
2
Open parameter menu.
3
Step through the menu 
to the PROBE param-
eter.
4
Move to the setting fi eld.
If the probe code displayed has only two digits present, then 
after step 4 select the slot required by inc/dec to one of the 
three available four digit slots. Return to the parameter fi eld by 
pressing SET and continue to PROBE D1 as normal step 5.
5
Step through the menu 
to the PROBE D1 pa-
rameter.
6
Move across to the set-
ting fi eld.
Preparation for operation